PERFORMANCE
| Accuracy of Reading: |
± .05°C |
| Warm-up Time: |
<10 minutes* |
| Recovery Time from Test: |
<2 seconds |
| Liquid Level |
500 ml |
| *Ambient @20°C |
SPECIFICATIONS
| Construction: |
All metal, chrome plated |
| Weight: |
5 lbs |
| Readout: |
Bright LED digital display |
| Sensor: |
Electronic Solid State |
| Power Input: |
120 VAC, 60 cycle
220 - 240 VAC, 50 cycle |
| Power Consumption: |
72 watts |
A Calibration Port is Provided for Insertion of a National Institute of Standards and Technology (NIST) Reference Mercury-In-Glass Thermometer
FEATURES
- Eight bit microprocessor control
- High stability internal voltage reference
- 12 bit digital accuracy from A/D converter
- Internal AC power transformer with AC connections isolated
- Visual indication of system operation during warm-up (dot scan)
- Visual indication of heater activitiy via dot display
- Visual and audible indication of system malfunction (Malfunction is displayed as an error number)
- RFI protected
SYSTEM SHUTDOWN UNDER THE FOLLOWING CONDITIONS
- No output from shorted or open temperature sensor
- No change in ambient temperature after 15 minutes
- Temperature of solution <33.8°C / >34.2°C
- RFI present
- No reading from internal A/D converter
ERROR STATUS CODES
Error 1 = Sensor Failure
Error 2 = RFI present
Error 3 = Temperature >34.2°C
Error 4 = No temperature change
Error 5 = Temperature <33.8°C
WARRANTY
One year, parts and labor.
|